Calibration of Turbidity Meter
1.0 PURPOSE
The purpose of the standard operating
procedure is to provide guidelines for calibration of
Turbidity
meter.
2.0 SCOPE
The scope of this standard operating procedure
includes calibration of Viscometer
(oscillatory).
3.0 RESPONSIBILITY
3.1 Section Incharge
3.2 Technical Manager
3.3 Q.A. Manager
4.0 PROCEDURE
4.1 Switch on the instrument.
4.2 Using the standard operating conditions start
the operations of the instrument.
4.3 Set the instrument at zero.
4.4 Pour the standard turbidity solution in the
glass tube / cuvette and measure the turbidity.
4.5 Ensure that the glass tube / cuvette is cleaned/washed
before and after use.
4.6 Make the measurements of turbidity in each
range of the instrument or in different ranges of
the
Instrument.
4.7 The turbidity shall not differ from the stated
values of the reference standards.
4.8 Correlate the observed values of turbidity
with the specified/stated values of the reference
standard
and make adjustments, if necessary.